An unusual foreign body in the breathing circuit


Department of Anaesthesia and Intensive Care, Gian Sagar Medical College and Hospital, Banur, Distt. Patiala, Punjab, India

Date of Web Publication3-Jun-2011

Correspondence Address:
Vikramjeet Arora
Department of Anaesthesia and Intensive Care, Gian Sagar Medical College and Hospital, Banur, Distt. Patiala, Punjab
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.81850

Rights and Permissions

How to cite this article:
Arora V. An unusual foreign body in the breathing circuit. J Anaesthesiol Clin Pharmacol 2011;27:294

How to cite this URL:
Arora V. An unusual foreign body in the breathing circuit. J Anaesthesiol Clin Pharmacol [serial online] 2011 [cited 2019 Aug 21];27:294. Available from: http://www.joacp.org/text.asp?2011/27/2/294/81850

Sir,

Failure in properly checking equipment is a factor in many critical incidents. Proper checking of equipment can help prevent equipment-related morbidity and mortality, improve preventive maintenance, and educate the anesthesia provider about equipment. Unfortunately, failure to perform a proper check before use is common. [1] We encountered an unusual foreign body in the breathing circuit and felt the need to re-emphasise the importance of the cockpit drill before induction of anesthesia.

There are numerous reports of obstruction of anesthesia breathing circuits by different foreign bodies. [2],[3] However, we did not come across literature reporting breathing circuit obstruction caused by a cap of urobag tubing. While checking the anesthesia machine and breathing circuit, it was noticed that intra-circuit pressure did not decrease on opening the end of elbow connector to the atmosphere. On thorough inspection, no source of obstruction could be found, but on turning the circuit upside down and removing the elbow connector, a foreign body, which was, in fact, a cap of catheter end of urobag tubing, was noticed. It was in the Y-piece of breathing circuit with the open end abutting the bifurcation and blind end in the straight limb of Y-piece [Figure 1]. On initial inspection, the foreign body appeared as a continuation of the elbow connector because of the same color. It took some time to find the exact site of obstruction, which could had been sufficient to cause hypoxia.
Figure 1: Urobag cap in the Y-piece not easily identifiable

Click here to view


The exact reason for this unusual occurrence could not be determined. It was the first case on that day; moreover, on the previous day, three cases were managed in the same operation theatre uneventfully using the same circuit. One hypothesis is that nontechnical staff of the operation theatre, while cleaning at the end of the previous day, might have misconnected the urobag cap thinking it to be a part of the breathing circuit. A foreign body lodged in the elbow connector (a glass ampoule fragment after it was used to open a propofol ampoule) causing an impossible ventilation has been reported earlier. [4]

Although, the exact cause of this unusual incident could not be concluded, but its timely detection definitely avoided a potential catastrophe. We recommend that readers thoroughly follow the pre-use checkout guidelines to prevent critical incidents and improve patient safety. Lastly, the nontechnical staff of the operation theatre should be educated and trained about isolation and disposal methods of the items such as cap of urobag tubing or similar items, which have the potential to cause circuit obstruction.

 
   References Top

1.Dorsch JA, Dorsch SE. Equipment checking and maintenance. In: Dorsch JA, Dorsch SE, editors. Understanding anaesthesia equipment. 5 th ed. New Delhi: Lippincott Williams and Wilkins, a Wolters Kluwer business; 2008. p. 931-54.  Back to cited text no. 1
    
2.Foreman MJ, Moyes DG. Anaesthetic breathing circuit obstruction due to blockage of tracheal tube connector by foreign body-two cases. Anaesth Intensive care 1999;27:73-5.  Back to cited text no. 2
[PUBMED]    
3.Krensavage TJ, Richards E. Sudden development of anaesthesia circuit obstruction by an end-tidal CO 2 cap in the gas sampling elbow. Anaesth Analg 1995;81:207.  Back to cited text no. 3
    
4.Gallacher BP, Kelly M, Mora RR. Failure to ventilate due to glass ampule fragment occlusion of the breathing circuit. Anaesthesiology 1997;87:180.  Back to cited text no. 4
    


    Figures

  [Figure 1]


This article has been cited by
1 An Audit of Level Two and Level Three Checks of Anaesthesia Delivery Systems Performed at Three Hospitals in South Australia
N. Sweeney,H. Owen,R. Fronsko,E. Hurlow
Anaesthesia and Intensive Care. 2012; 40(6): 1040
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1293    
    Printed63    
    Emailed2    
    PDF Downloaded241    
    Comments [Add]    
    Cited by others 1    

Recommend this journal