Leaks in the vaporizer unit: Still a possibility


Department of Anaesthesiology, Sri Devaraj Urs Medical College, Tamaka, Kolar, India

Date of Web Publication11-Aug-2011

Correspondence Address:
B R Krishna Kumar
24/1, 6th main, 6th cross, postal colony, Sanjaynagar, Bangalore
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.83700

Rights and Permissions

How to cite this article:
Krishna Kumar B R, Anand T T, Dinesh K, Sindhu B S. Leaks in the vaporizer unit: Still a possibility. J Anaesthesiol Clin Pharmacol 2011;27:415

How to cite this URL:
Krishna Kumar B R, Anand T T, Dinesh K, Sindhu B S. Leaks in the vaporizer unit: Still a possibility. J Anaesthesiol Clin Pharmacol [serial online] 2011 [cited 2019 May 21];27:415. Available from: http://www.joacp.org/text.asp?2011/27/3/415/83700

Sir,

Modern day vaporizers are reliable and come with many safety features. We report a case of an unusual leak from the vaporizer unit. The incident occurred while conducting anesthesia in a patient with mandibular fracture, posted for open reduction and internal fixation, on replacement of the halothane vaporizer with isoflurane vaporizer. After replacement, there was sudden under-filling of the reservoir bag after turning the control dial on. The flow meters showed no change in the flow, which meant there was a leak. The connections within the circuit, between the circuit, and the common gas outlet and filler cap of the vaporizer were checked out for leaks, but no leaks were detected. On checking the vaporizer mount, a needle was found between mounting system and the vaporizer unit, which was causing the leak. The leak was then rectified.

Detachable mounting systems, like Selectatec, permit movement and change of the vaporizer unit. These consist of a pair of port valves for the vaporizer. Each vaporizer has a special mounting bracket that fits over the nipples. The weight of the vaporizer and an O-ring around each port valve create a seal between the mounting system and the vaporizer. [1]

In case of a leak in the vaporizer, the anesthesia delivery machine can function normally if the vaporizer remains switched off. On switching on the vaporizer, the fresh gas flow is reduced with little or no vapor. A common cause of leak is failure to replace the filler cap or tighten it. [2] If an incorrect cap is used on a filler device, leak may occur. [3] Another common cause of the leak is the absent or damaged O-ring. [1],[4] Other locations for leaks include the selector valve, mounting mechanism, interlock device, vaporizer inlet, or outlet connections.

In our case, though a needle was present between the mounting system and vaporizer [Figure 1], it was possible to lock the locking lever and the control dial could be turned on. This led to leakage of fresh gas flow and under-filling of the reservoir bag. Consequences of leaks include alteration in the flow and composition of inspiratory gas and pollution of the operating room air. The former can cause hypoxia, re-breathing or awareness in a patient if not identified and rectified.
Figure 1: Needle causing the leak

Click here to view


In conclusion, detachable mounts are convenient but are a potential source of leak. We recommend that utmost care be taken during dismounting and remounting of vaporizers to avoid damage to them and prevent leaks.

 
   References Top

1.Dorsch JA, Dorsch SE. Understanding anesthesia equipment. 5 th ed. Philadelphia: Williams and Wilkins; 2008. p. 181.  Back to cited text no. 1
    
2.Meister GC, Becker KE Jr. Potential fresh gas flow leak through Drager Vapor 19.1 vaporizer with key index fill port. Anaesthesiology 1993; 78: 211-2.  Back to cited text no. 2
    
3.Ngan J, Cashen D, Nicols G. Sevoflurane vaporizers. Can J Anesth 1999; 46: 200.  Back to cited text no. 3
    
4.James RH. Defective Selectatec O-rings. Anaesthesia 1995; 50:184-5.  Back to cited text no. 4
[PUBMED]    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1221    
    Printed71    
    Emailed0    
    PDF Downloaded238    
    Comments [Add]    

Recommend this journal