Failed insertion of endotracheal tube through classic laryngeal mask airway


Department of Anaesthesiology and Intensive Care, Rajendra Prasad Institute of Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India

Date of Web Publication11-Aug-2011

Correspondence Address:
Renu Sinha
S-6, First Floor O.P.D. Block, R.P. Center, All India Institute of Medical Sciences, Ansari Nagar, New Delhi-110029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.83708

Rights and Permissions

How to cite this article:
Sinha R, Ray BR, Dey D, Swetha S. Failed insertion of endotracheal tube through classic laryngeal mask airway. J Anaesthesiol Clin Pharmacol 2011;27:423

How to cite this URL:
Sinha R, Ray BR, Dey D, Swetha S. Failed insertion of endotracheal tube through classic laryngeal mask airway. J Anaesthesiol Clin Pharmacol [serial online] 2011 [cited 2019 Jun 25];27:423. Available from: http://www.joacp.org/text.asp?2011/27/3/423/83708

Sir,

The use of classic laryngeal mask airway (cLMA) as a conduit for intubation is a known technique in pediatric difficult airway. We faced a unique problem during the management of an unanticipated difficult airway in a neonate while negotiating the endotracheal tube (ETT) through cLMA.

A 1-month-old infant, weighing 3.8 kg, had an unanticipated difficult airway with failure to intubate the trachea despite repeated direct laryngoscopic attempts. Adequate ventilation was achieved with cLMA size 1.0. A flexible fiberoptic bronchoscope (FOB; size 2.8) through cLMA showed grade 4 glottic view. [1] It was decided that a size 3.5 uncuffed ETT to be passed over the FOB and tracheal intubation attempted using cLMA as a conduit. Despite successfully negotiating the FOB into the trachea, we were unable to pass the 3.5- mm ETT through the cLMA 1.0. Subsequently, tracheal intubation was achieved with a size 3.0-mm ID uncuffed ETT through the same cLMA 1.0 by using the blind technique.

After removal of cLMA, we failed to pass the 3.5-mm ETT through the cLMA even with force. Later, while trying to negotiate 3.5-mm ETT through all (four in daily use) available cLMAs of size 1.0, we were able to pass the 3.5-mm ETT through only one cLMA. In other cLMAs, the ETT stuck in the shaft of the LMA after crossing the connector. The 3.5-mm ETT also passed through a freshly opened size 1.0 cLMA.

The size 3.5-mm ID uncuffed ETT should pass through size 1.0 cLMA, according to manufacturer's recommendation. There is no mention of the standard internal diameter of the shaft of LMAs in the available literature. Although manufacturers recommend that LMA be used up to 40 times after autoclaving, there have been reports of successful use of LMA up to 100 times. [2] In our institute, we use cLMA more than the recommended 40 times after testing. In the case reported, three of four LMAs, including the one that allowed passage of the ETT, had been used more than 40 times. The 3.5 mm ETT could not be passed through a cLMA, which had been used less than 40 times. It is not known whether autoclaving produces changes in the outer and inner diameter of shaft of LMA.

We recommend that before using cLMA as a conduit for intubation, the maximum size of tracheal tube that can be passed through a particular LMA should be checked before, to avoid unanticipated difficulties. Manufacturer's details about the internal diameter may be helpful for further evaluation of change in the diameter of shaft after autoclaving.

 
   References Top

1.Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of the laryngeal mask airway position. Anesth Analg 1993;76:457.  Back to cited text no. 1
[PUBMED]    
2.Goodman EJ, Christenson E, Douglas AM, Ziegler EJ, Lewis BR. Reusable laryngeal mask airways can be used more than 40 times. J Clin Anesth 2008;20:109-15.  Back to cited text no. 2
[PUBMED]  [FULLTEXT]  




 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References

 Article Access Statistics
    Viewed983    
    Printed61    
    Emailed0    
    PDF Downloaded220    
    Comments [Add]    

Recommend this journal