Move the anesthesia workstation cautiously!


Department of Anesthesiology & Critical Care Medicine, Indira Gandhi Institute of Medical Sciences, Patna, Bihar, India

Date of Web Publication24-Jan-2014

Correspondence Address:
Prakash K Dubey
Eľ, Indira Gandhi Institute of Medical Sciences Campus, Sheikhpura, Patna - 800 014, Bihar
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.125735

Rights and Permissions

How to cite this article:
Dubey PK. Move the anesthesia workstation cautiously!. J Anaesthesiol Clin Pharmacol 2014;30:121-2

How to cite this URL:
Dubey PK. Move the anesthesia workstation cautiously!. J Anaesthesiol Clin Pharmacol [serial online] 2014 [cited 2019 Sep 21];30:121-2. Available from: http://www.joacp.org/text.asp?2014/30/1/121/125735

Sir,

Inadvertent damage to lines and cords during repositioning of anesthesia workstation after a patient is anesthetized can occur due to carelessness. However, damage to the blood pressure tubing or the sample line tubing of capnograph can affect the anesthesia monitoring significantly. We present a case of erroneous drop in end tidal carbon dioxide level due to damage of the sampling tube during general anesthesia.

A 65 years American Society of Anesthesiologists II male of 51kg body weight was posted for anterolateral decompression of caries' spine under general anesthesia. Anesthesia was induced with morphine 5mg and thiopentone 300mg intravenous after instituting monitoring in the form of electrocardiography, non-invasive blood pressure, SpO 2 and EtCO 2 . Tracheal intubation was facilitated with vecuronium 6mg and the trachea intubated with an 8.0 mm diameter tracheal tube with ease. Anesthesia was maintained with 0.6% isoflurane and 60% nitrous oxide in oxygen along with intermittent positive pressure ventilation (IPPV).

The patient was repositioned in right lateral decubitus position to facilitate surgical exposure. During this process, the anesthesia workstation had to be moved to make room for the OR personnel. Routine auscultation of the chest was done to rule out any displacement of the tube after repositioning and was found satisfactory.

Suddenly it was noticed that the capnogram had changed into a dual waveform [Figure 1]. The EtCO 2 level had also dropped from 34 mmHg to 20 mmHg, the inspiratory and end tidal values of isoflurane changed from 0.5% and 0.4% respectively to 0.3% and 0.3% and that of nitrous oxide from 54% and 53% to 28% and 35% respectively. There was, however, no change in oxygen saturation or hemodynamic parameters. Chest auscultation was again performed to ensure bilateral equal air entry. Integrity of the sample line right from the sidestream connector was checked that revealed a damaged sample line. The sample line was immediately replaced with a fresh one and the EtCO 2 value returned to the baseline. The capnogram now displayed a normal waveform. Surgery and anesthesia continued and terminated uneventfully.
Figure 1: Dual waveform capnogram showing an EtCO2 level of 20 mm of Hg

Click here to view


Important causes for a dual wave form capnogram include mechanical factors like sampling leaks, loose connection between sampling tubes and the monitor, and breaks in the water filter of the carbondioxide analyzer. Pathological causes include severe kyphoscoliosis and differential emptying of the lungs. [1] Physiological condition like pregnancy and obesity can cause an upswing at the end of Phase III. [2] Iatrogenic causes include endobronchial intubation, which was ruled out in our case.

Mechanics of change in capnogram with a "tail-up" due to damage to the sample line has been investigated by Tripathi and Pandey. [3] During IPPV, pressure in the breathing circuit is lower during exhalation, allowing air to enter through the hole in the sampling tube causing a low EtCO 2 . The authors hypothesized that with inspiration, the positive pressure in the breathing circuit, transmitted to the sampling tube, prevents air admixture and the upsurge in carbon dioxide that is being displayed, giving the capnograph a"tailsup" appearance.

Isolated drop in EtCO 2 value might have confused us about the possible cause of this change. Typical change in the waveform prompted us to search for a leak. During movement of the workstation, the sample line dangling on the floor was crushed inadvertently as the capnogram was normal before this. Immediately the waveform changed and the level of inhalational agents and EtCO 2 went down. These changes were reproducible by changing the new sample line with the damaged one.

Patient repositioning during general anesthesia is known to potentially cause injury. Repositioning of workstation may also jeopardize patient safety, as illustrated by this case. This case also emphasizes the significance of monitoring and alarm system during equipment movement under anesthesia. Alarms are often caused by manipulations performed by the medical staff working with the anesthesia workstation and may be annotated as "not relevant,"changing the significance of the alarm. [4] Also, awareness of this possible cause may help the clinician in narrowing down the reason of sudden decrease in EtCO 2 value due to such a mishap.

Extreme precaution during movement of the work station is needed to avoid such occurrences as a sudden decrease in EtCO 2 value can lead to confusion especially during procedures like neurosurgery or where carbondioxide is used as insufflation gas.

 
  References Top

1.Nichols KP, Benumof JL. Biphasic carbon dioxide excretion waveform from a patient with severe kyphoscoliosis. Anesthesiology1989;71:986-7.  Back to cited text no. 1
[PUBMED]    
2.Dorsch JA, Dorsch SE. Gas monitoring. Understanding Anesthesia Equipment. 5 th ed. Philadelphia: Wolters Kluwer / Lippincott Williams & Wilkins; 2008. p. 685-727.  Back to cited text no. 2
    
3.Tripathi M, Pandey M. Atypical "tails-up" capnograph due to breach in the sampling tube of side-stream capnometer. J Clin Monit Comput 2000;16:17-20.  Back to cited text no. 3
[PUBMED]    
4.Schmid F, Goepfert MS, Kuhnt D, Eichhorn V, Diedrichs S, Reichenspurner H, et al. The wolf is crying in the operating room: Patient monitor and anesthesia work station alarming patterns during cardiac surgery. Anesth Analg 2011;112:78-83.  Back to cited text no. 4
[PUBMED]    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed1057    
    Printed30    
    Emailed1    
    PDF Downloaded217    
    Comments [Add]    

Recommend this journal