Leak in anesthesia workstation: An unusual cause


Department of Anesthesiology and Critical Care, Deen Dayal Upadhyay Hospital, New Delhi, India

Date of Web Publication15-Mar-2018

Correspondence Address:
Ajay Kumar
BB/49-C, Janakpuri, New Delhi - 110 058
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


Rights and PermissionsRights and Permissions

How to cite this article:
Aggarwal R, Kumar A. Leak in anesthesia workstation: An unusual cause. J Anaesthesiol Clin Pharmacol 2018;34:135-6

How to cite this URL:
Aggarwal R, Kumar A. Leak in anesthesia workstation: An unusual cause. J Anaesthesiol Clin Pharmacol [serial online] 2018 [cited 2019 Oct 18];34:135-6. Available from: http://www.joacp.org/text.asp?2018/34/1/135/227582



We would like to report an unusual cause of leak in the anesthesia workstation. Dräger Fabius ® GS anesthesia workstation was checked according to manufacturer's recommendations.[1] A 50-year-old female was administered general anesthesia for laparoscopic cholecystectomy using standard protocol. Her trachea was intubated (size 7 cuffed oro-tracheal tube) and lungs were manually ventilated with oxygen and nitrous oxide in halothane using a closed circuit. Sevoflurane vaporizer (Dräger Vapor ® 2000) was arranged and installed on the workstation; we switched to using sevoflurane. Immediately, there was under-filling of the reservoir bag. We increased the fresh gas flows and closed the adjustable pressure limiting valve completely. 100% oxygen in sevoflurane was administered. All connections in the breathing system and tracheal tube cuff were checked for leak. The sevoflurane vaporizer was turned off; but there was no improvement. The breathing system was changed to Bain's circuit. This increased the difficulty in ventilation, requiring the use of oxygen flush intermittently. We switched back to closed circuit; with it, we could ventilate using increased flows. Patient did not desaturate anytime and her vitals remained stable. Alternative anesthesia workstation was arranged and used for the rest of surgery which was uneventful.

The “leaking workstation” was put on leak and compliance self-test. It showed leak >350 ml. On starting oxygen flow there was under filling of the test lung. Workstation had no sign of external damage. Both halothane and sevoflurane vaporizers were stable on the manifold, at same level, with no tilting. Savlon-soaked gauze was dabbed on the backbar. Continuous bubbling was seen at the site of attachment of sevoflurane vaporizer to the backbar, implying leak [Figure 1]. Upon turning-on the vaporizer, odor of sevoflurane could be appreciated. The sevoflurane vaporizer was removed from the backbar. Glass pieces were seen between the vaporizer and the mounting system [Figure 2]. These were removed, vaporizer was re-mounted and the workstation rechecked; there was no leak.
Figure 1: Continuous bubble formation depicting leak

Click here to view
Figure 2: Glass pieces on the mounting system

Click here to view


The vaporizer is fitted onto the port valves on the backbar. The weight of the vaporizer and the O-ring around each port valve create a seal between the mounting system and the vaporizer.[2] Fresh gas flow leak due to malpositioning of the vaporizer has been reported.[3],[4] In this case, broken pieces of glass prevented a tight seal around the port valves of the vaporizer causing leak in fresh gas flow even when vaporizer was not turned on. Perhaps, a glass ampoule kept on top of the vaporizer or else the roof of anesthesia workstation had fallen and got caught between the sevoflurane vaporizer and the mounting system during its installation.

Resting half-filled, empty or spare ampoules on the vaporizer top or workstation roof is not an uncommon practice; this should be discouraged. Spare ampoules should be neatly placed in their racks and used ampoules should either be discarded or kept in dedicated racks. Vaporizer should be installed on the backbar with utmost care, followed by workstation safety check. Practice of replacing or mounting a vaporizer while the workstation is in use should be deterred. It is not possible to recheck the anesthesia workstation at this stage and safety norms are flouted in doing so.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Dräger medical: Fabius GS Operator's Manual. 3rd ed. Germany: Dräger Medical AG & Co. KG; 2006. Appendix ‒ Daily and preuse checkout form.   Back to cited text no. 1
    
2.
Dorsch JA, Dorsch SE. Understanding anesthesia equipment. 5th ed. Philadelphia, USA: Lippincott Williams & Wilkins; 2008. p. 122-90.  Back to cited text no. 2
    
3.
Kim HJ, Kim MW. Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer – A case report. Korean J Anesthesiol 2010;59:270-4.  Back to cited text no. 3
    
4.
Krishna KB, Anand T, Dinesh K, Sindhu B. Leaks in the vaporizer unit: Still a possibility. J Anaesthesiol Clin Pharmacol 2011;27:415.  Back to cited text no. 4
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed541    
    Printed15    
    Emailed0    
    PDF Downloaded91    
    Comments [Add]    

Recommend this journal