Water contamination of the central air supply: Out of sight, out of mind!


Department of Anaesthesia, Chacha Nehru Bal Chikitsalaya, New Delhi, India

Date of Web Publication15-Mar-2018

Correspondence Address:
Anju Gupta
A-437, Sarita Vihar, New Delhi - 110 076
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_175_16

Rights and Permissions

How to cite this article:
Kamal G, Gupta A, Bathla S, Gupta A. Water contamination of the central air supply: Out of sight, out of mind!. J Anaesthesiol Clin Pharmacol 2018;34:139-40

How to cite this URL:
Kamal G, Gupta A, Bathla S, Gupta A. Water contamination of the central air supply: Out of sight, out of mind!. J Anaesthesiol Clin Pharmacol [serial online] 2018 [cited 2019 Oct 13];34:139-40. Available from: http://www.joacp.org/text.asp?2018/34/1/139/227560



Water in flowmeters may cause bobbin malfunction, deliver inaccurate flows, and lead to hypoxia.[1],[2],[3] Remote location of medical gas pipeline system (MGPS) can deter anesthesiologist from being routinely involved in its day-to-day maintenance and thus, being caught unaware in the case of malfunction. We faced a harrowing incident wherein sudden influx of water column in air flow meter of anesthesia workstation (Datex Ohmeda Aestiva 5) was noticed due to faulty compressed air delivery system.

A 35-week-old preterm neonate was posted for emergency laparotomy under general anesthesia for duodenal atresia. The anesthesia workstation in our emergency OT passed the pre-use check. After rapid sequence induction, the baby was intubated with 3.5 mm I.D. endotracheal tube and control knob of air flow meter was opened for the maintenance of anesthesia. To our great surprise, there was a sudden influx of water in the air flow meter from the needle valve and its level was rising fast. The bobbin got stuck at the bottom and stopped rotating. The circuit was disconnected and ventilation commenced with bag mask device. The pipeline was disconnected from the anesthesia workstation, and sevoflurane in 100% oxygen (from cylinder in the standby anesthesia machine) was used for anesthesia maintenance.

Water is a usual contaminant in a compressed air system, and its amount may air vary from 2.5 g/m3 to over 40 g/m3 depending on climatic condition.[1] Air gets heated during the compression process and condensed moisture during cooling is removed in the after cooler. However, about 20 g/m3 may still remain in the compressed air unless removed by a dryer.[1] The probable causes of water contaminating the MGPS include manual water drain operated at longer time interval or drain clogging, absence of desiccant air dryer, exposure of central air pipeline to atmosphere and absence of alarms for water level.[2],[3],[4],[5]

Fortunately, we noticed the water column in air flow meter immediately before it could contaminate the fresh gas flow. Detailed investigation revealed that there were reports of ventilator malfunction in the Pediatric Intensive Care Unit earlier due to faulty automatic drain of the compressed air system which was replaced. Subsequently, following a recent ventilator malfunction, fault in desiccant dryer was detected and repaired. Probably, this time purging of some of the “risers and branch lines” was not done. Hence, the water accumulated in the branch line (of emergency OT) as a result of dryer malfunction and exposure of air pipeline to the humid atmosphere (due to rainy season) during the repair of air dryer were probably the main culprits.

All components of MGPS were checked by the service engineer; the pipeline was flushed thoroughly and to guard against such events in future checkups for medical air quality of pipeline were made more frequent. Technical engineer of the anesthesia workstation also confirmed the presence of water in the air flow meter tubing and dried all the channels properly.

In conclusion, maintenance of MGPS should be directly supervised by anesthetists and medical air quality should be checked after maintenance work before its reconnection to the anesthesia machines after though purging.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Dorsch JA, Dorsch SE, editors. Medical gas pipeline systems. In: Understanding Anaesthesia Equipment. 5th ed. Philadelphia, USA: Lippincott Williams and Wilkins; 2008. p. 25-50.  Back to cited text no. 1
    
2.
Anand LK, Kapoor D, Kazal S. Water in the flowmeters: Still a possibility in the modern era! Anesth Analg 2013;117:281-3.  Back to cited text no. 2
    
3.
Okuyama M, Nakamura I, Kemmotsu O. Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia. Masui 2000;49:652-4.  Back to cited text no. 3
    
4.
Manjuladevi M, Upadhyaya KV, Sathyanarayana P, Rani R, Shah V. Critical incident is a possibility with water in flowmeter. Indian J Anaesth 2014;58:760-2.  Back to cited text no. 4
[PUBMED]  [Full text]  
5.
Hay H. Contamination of piped medical gas supply with water. Eur J Anaesthesiol 2000;17:512-4.  Back to cited text no. 5
    




 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References

 Article Access Statistics
    Viewed574    
    Printed10    
    Emailed0    
    PDF Downloaded87    
    Comments [Add]    

Recommend this journal