Users Online: 880 Home Print this page Email this page Small font sizeDefault font sizeIncrease font size  
Home | About us | Editorial board | Search | Ahead of print | Current issue | Archives | Submit article | Instructions | Subscribe | Contacts | Login 


RSACP wishes to inform that it shall be discontinuing the dispatch of print copy of JOACP to it's Life members. The print copy of JOACP will be posted only to those life members who send us a written confirmation for continuation of print copy.
Kindly email your affirmation for print copies to dranjugrewal@gmail.com preferably by 30th June 2019.

 

 
Table of Contents
LETTER TO EDITOR
Year : 2018  |  Volume : 34  |  Issue : 2  |  Page : 257-258

Inappropriately sized connector: An ingredient for catastrophe!?


Department of Anaesthesiology and Critical Care, All India Institute of Medical Sciences, Jodhpur, Rajasthan, India

Date of Web Publication16-Jul-2018

Correspondence Address:
Rashmi Syal
Department of Anaesthesiology and Critical Care, All India Institute of Medical Sciences, Jodhpur, Rajasthan
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_252_17

Rights and Permissions

How to cite this article:
Syal R, Chhabra S, Mohammed S, Bhatia P. Inappropriately sized connector: An ingredient for catastrophe!?. J Anaesthesiol Clin Pharmacol 2018;34:257-8

How to cite this URL:
Syal R, Chhabra S, Mohammed S, Bhatia P. Inappropriately sized connector: An ingredient for catastrophe!?. J Anaesthesiol Clin Pharmacol [serial online] 2018 [cited 2019 Nov 22];34:257-8. Available from: http://www.joacp.org/text.asp?2018/34/2/257/236666



Madam,

Pre-use check of endotracheal tube (ETT) is a routine anesthesia practice. Despite this, manufacturing defects are frequently encountered which can go unnoticed on visual inspection and may lead to catastrophic situations.[1]

We report one such case of a 54-year-old female (weight 75 kg) posted for laparoscopic cholecystectomy. Following the induction of general anesthesia as per local standards, patient was intubated with 7.5 mm Internal Diameter (ID) ETT (Sterimed Health Care Technologies, LLC, Florida, USA, manufactured at Sterimed Medical Devices Ltd. Bahadurgarh, Haryana). While connecting the ETT to the circuit of anesthesia machine, we found that the connection was loose and circuit was being easily disconnected from the tube. The connection was held manually to ventilate the patient. In the meantime, a properly fitting new connector of same size ETT (different manufacturer) was arranged. Since, the faulty connector could not be removed from the tube; it was not possible to replace it with the new connector. To prevent frequent intraoperative disconnections, we applied a layer of micropore adhesive over the connector [Figure 1]. The connector now fitted snugly into the circuit and remained connected during rest of the surgery.
Figure 1: A layer of micropore adhesive applied to the endotracheal tube connector

Click here to view


Postoperatively, the connector was measured using a digital Vernier caliper and was found to be of smaller size (14.76 mm) in contrast to 15 mm standard connector [Figure 2]. The difference was small, but clinically significant.
Figure 2: Measurement of the connector in question (a) and a standard connector (b) using digital vernier caliper

Click here to view


There are many reports in literature where manufacturing defects involving the cuff inflation system or patency of connectors have been described.[2],[3],[4],[5] All these reports emphasize on pre-use check for cuff leak and tube patency. The routine check of airway connectors for circuit compatibility has not been recommended as their size has been standardized (15 mm diameter) so that all airway devices are readily connectable to anesthesia circuit. The manufacturing defect in connectors can prove to be catastrophic especially in difficult airway scenarios where replacement of ETT may be challenging. Such incidents raise a concern regarding need for more stringent quality checks.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Gettelman TA, Morris GN. Endotracheal tube failure: Undetected by routine testing. Anesth Analg 1995;81:1313.  Back to cited text no. 1
    
2.
Sofi K, El-Gammal K. Endotracheal tube defects: Hidden causes of airway obstruction. Saudi J Anesth 2010;4:108-10.  Back to cited text no. 2
    
3.
Shamshery C, Kannaujia AK, Gautam S. Ventilation failure due to endotracheal tube T-connector defect. Indian J Anaesth 2010;54:357-8.  Back to cited text no. 3
[PUBMED]  [Full text]  
4.
Bharti N, Bala I, Sharma K. Endotracheal tube connector defect as a cause of high airway pressure. Paediatr Anaesth 2012,22:502-3.  Back to cited text no. 4
    
5.
Jain D, Bala I. Manufacturing defect of endotracheal tube connector: A cause of airway obstruction. J Anaesthesiol Clin Pharmacol 2014;30:583-4.  Back to cited text no. 5
[PUBMED]  [Full text]  


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed524    
    Printed12    
    Emailed0    
    PDF Downloaded0    
    Comments [Add]    

Recommend this journal