An incidental finding of endotracheal tube obstruction at the level where inflation line enters into the tube


Department of Critical Care and Anaesthesia, Medanta The Medicity Hospital, Gurgaon, Haryana, India

Date of Web Publication11-Oct-2018

Correspondence Address:
Arvind Kumar
Department of Critical Care and Anaesthesia, Medanta The Medicity Hospital, Sector 38 Gurgaon, Haryana
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_207_15

Rights and Permissions

How to cite this article:
Kumar A, Khanna S, Mehta Y. An incidental finding of endotracheal tube obstruction at the level where inflation line enters into the tube. J Anaesthesiol Clin Pharmacol 2018;34:417-8

How to cite this URL:
Kumar A, Khanna S, Mehta Y. An incidental finding of endotracheal tube obstruction at the level where inflation line enters into the tube. J Anaesthesiol Clin Pharmacol [serial online] 2018 [cited 2018 Dec 19];34:417-8. Available from: http://www.joacp.org/text.asp?2018/34/3/417/243156



Madam,

A 33-year-old female, ASA physical status I patient, presented for subtotal thyroidectomy. All standard monitors including electrocardiography (ECG), pulse oximetery (SPO2), and noninvasive blood pressure (NIBP) were attached. Standard anesthesia induction and endotracheal intubation with 7.5-mm internal diameter (ID) cuffed armoured (Mallinkrodt) endotracheal tube (ETT) was done. Lungs were ventilated using volume control mode after connecting to the anesthesia machine with close circuit. Bilateral air entry, lung compliance, and endtidal carbon dioxide (ETCO2) were confirmed. In this case, the head end of the patient was away from the anesthesia machine, hence, long length closed circuit was used. After approximately 1 h and 30 min of commencement of the surgery, suddenly monitor stopped displaying ETCO2, airway pressure rose above 35 cm of water (cmH2O) which was earlier 16–17 cmH2O, and endtidal volume was zero. While SPO2 remained 100%, we checked all possible causes sequentially. The anesthesia machine, circuit, all valves, and connections were checked. On manual ventilation, high resistance was felt and there was significantly decreased bilateral air-entry on auscultation. Suction catheter was inserted through the ETT to remove possible mucous plug/blood clot but it could not be negotiated beyond 2–3 cm. A check laryngoscopy was done and ETT was changed. Now, manual ventilation was optimal and ETCO2 curve was normal. The surgery continued and trachea was extubated after surgery uneventfully.

Later, it was found that there was an obstruction at the level where the pilot balloon tube was entering into the ETT [Figure 1]. It was only evident when the cuff was inflated, so to confirm, we injected methylene blue into the pilot balloon [Figure 2].
Figure 1: Inside view of intraluminal obstruction

Click here to view
Figure 2: External view after methylene blue injection

Click here to view


Many manufacturing defects in ETT have been observed in the past, which include overpressure cuff herniation,[1],[2] intraluminal tracheal obstruction,[2] distal end intraluminal meniscal obstruction,[3] and tube kink.[4] Patients with controlled ventilation, close observation of changes in peak airway pressures, lung compliance, ETCO2, and endtidal volume are warranted.[5]

In this case, occlusion was due to a defect in the inner film of the tube at the level where pilot balloon's inflation line enters the wall of the ETT. Remarkably, when the cuff was deflated, the inner film also got settled on the wall and the lumen became normal in contour. Because it was a thyroid surgery, injury to the pleura or airway could not be overlooked and the surgeons were asked for any notable pleural or airway injury. Although there was an evidence of bilateral decreased air entry, when the suction catheter could not be negotiated through the tube, it was more in favor of endotracheal blockade.[2] However, a chance of kinking was remote because we used an armoured tube.[5]

In this case, the SpO2 always remained 100% but ETCO2 trace was not seen and there was high resistance on manual ventilation. All these factors suggested that some oxygen (O2) was being delivered with positive pressure but there was no endtidal trace. Because the endotracheal tube was under the drapes it was difficult to rule out tube handling, continuous pressure over the pilot balloon, or tube kinking, which could lead to this defect in this case. Oxygen and air were used for maintenance of anesthesia. The availability of cheap disposable products in the market today has resulted in the quality of the products being not up to standards.

Thus, this case highlights that there can be an obstruction due to a manufacturing defect and it can occur at any course of time during anesthesia. Vigilant monitoring and close observation of any change in ventilatory parameters can provide an early warning.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Davis Dan MD; Murphy Joseph MD; Pop Radu B.MS; Szmuk Peter MD. Airway obstruction due to endotracheal tube's lumen collapse secondary to cuff. Anesthesia and analgesia. 2011;112:1511-1512  Back to cited text no. 1
    
2.
Fameo CE. A not so apparent cause of intraluminal tracheal tube obstruction. Anesthesiology 1983;58:593  Back to cited text no. 2
    
3.
Khalid Sofi, Kariman EI-Gammal. Endotracheal tube defect: Hidden causes of airway obstruction. Saudi J Anaesth. 2010; 4:108-110.  Back to cited text no. 3
    
4.
Johnson KM, Lehman RE. Acute management of the obstructed endotracheal tube. Respiratory Care. 2012 Aug; 57:1342-4.  Back to cited text no. 4
    
5.
Chua WL, Ng AS. A defective endotracheal tube. Singapore Med J. 2002;43: 476-8  Back to cited text no. 5
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed109    
    Printed4    
    Emailed0    
    PDF Downloaded15    
    Comments [Add]    

Recommend this journal