Intubating laryngeal mask airway with transplanted pilot balloon?


1 Department of Anaesthesiology, PGIMER and RMLH, New Delhi, India
2 Specialist, Government of NCT, Pandit Madan Mohan Malviya Hospital, Delhi, India

Date of Web Publication31-Jan-2012

Correspondence Address:
Anju Gupta
437, Pocket A, Sarita Vihar, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.92469

Rights and Permissions

How to cite this article:
Bala V, Gupta A, Gupta N, Pawar M. Intubating laryngeal mask airway with transplanted pilot balloon?. J Anaesthesiol Clin Pharmacol 2012;28:131-2

How to cite this URL:
Bala V, Gupta A, Gupta N, Pawar M. Intubating laryngeal mask airway with transplanted pilot balloon?. J Anaesthesiol Clin Pharmacol [serial online] 2012 [cited 2021 May 9];28:131-2. Available from: https://www.joacp.org/text.asp?2012/28/1/131/92469

Sir,

Intubating laryngeal mask airway (ILMA) is often used in elective surgical procedures and emergency difficult airway management. It is made up of silicon rubber for its longevity. It is expected to perform best in first 40 uses; however, damage of device is not guaranteed up to 40 uses. The pilot balloon can get damaged while handling, washing, or cleaning. In such a case, it has to be discarded immediately. [1]

Recently, while doing the preuse check of an ILMA, we discovered that the pilot balloon assembly was disrupted. Since, it was a new device (hardly used twice) and as it is expensive, we did not discard it. We decided to replace the disrupted pilot balloon assembly of ILMA with the pilot balloon along with 2 cm inflation tubing of size 7 mm ID tracheal tube. The cut end of the inflation tubing along with pilot balloon fitted snugly into the inflation channel of ILMA is shown in [Figure 1]. The ILMA cuff could be inflated with the transplanted pilot balloon. The cuff pressures were sustained, as seen by the cuff dimensions and subsequently confirmed with cuff pressure manometer. The pilot balloon was subsequently fixed with the help of adhesive (Fevikwik, Pidilite Industries Limited, Mumbai, India)to prevent accidental leakage.
Figure 1: ILMA with damaged and transplanted pilot balloon

Click here to view


This device has been subsequently used more than 20 times and no problem was noted. The cuff and the airway tube of the repaired ILMA are cleaned in warm water with Endozime®. Thereafter, the inside of the airway tube using is cleaned with a soft bristle brush and then the device was rinsed thoroughly in warm running tap water.

We have described a novel indigenous technique for repair of the costly equipment. This technique can be utilized for repair of costly ILMA, but we should be careful and always check such a repaired device before use to prevent mishap.

 
  References Top

1.Singh M, Bharti R, Kapoor D. Repair of damaged supraglottic airway devices: A novel method. Scand J Trauma Resusc Emerg Med 2010;18:33.  Back to cited text no. 1
[PUBMED]  [FULLTEXT]  


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed1600    
    Printed64    
    Emailed0    
    PDF Downloaded244    
    Comments [Add]    

Recommend this journal