LETTER TO EDITOR |
|
Year : 2012 | Volume
: 28
| Issue : 2 | Page : 263-264 |
|
The i-gel™ - A promising airway device for magnetic resonance imaging suite
Susheela Taxak, Mamta Bhardwaj, Ajith Gopinath
Department of Anaesthesia and Critical Care, PGIMS, Rohtak, Amrita Institute of Medical Sciences, Kochi, India
Date of Web Publication | 11-Apr-2012 |
Correspondence Address: Mamta Bhardwaj 134, Eight Marla Colony, Patel Nagar, Hisar - 125 003, Haryana India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0970-9185.94917
How to cite this article: Taxak S, Bhardwaj M, Gopinath A. The i-gel™ - A promising airway device for magnetic resonance imaging suite. J Anaesthesiol Clin Pharmacol 2012;28:263-4 |
How to cite this URL: Taxak S, Bhardwaj M, Gopinath A. The i-gel™ - A promising airway device for magnetic resonance imaging suite. J Anaesthesiol Clin Pharmacol [serial online] 2012 [cited 2021 Mar 4];28:263-4. Available from: https://www.joacp.org/text.asp?2012/28/2/263/94917 |
Sir,
Anesthesia for magnetic resonance imaging (MRI) in pediatric patients is a challenging task. Most children require deep sedation as the procedure is lengthy, requires precise positioning, and patients' cooperation. These patients may require general anesthesia and securing of airway with endotracheal tube or supraglottic device. Various supraglottic devices have been described for use in MRI. The classic laryngeal mask (LMA-C TM ) is a helpful device in maintaining an adequate airway during MRI. [1],[2] With the exception of MRI-safe laryngeal mask airway (LMA), all LMAs have a variable quantity of ferromagnetic material that could reduce image quality. Recently, i-gel TM (Intersurgical, Workingham, UK), a new supraglottic airway device with noninflatable cuff, was introduced in pediatric sizes 1, 1.5, 2, and 2.5. We describe the use of i-gel TM in MRI suite in two children undergoing MRI for brain.
A 4-kg, 3-month-old child with seizures was scheduled to undergo MRI brain. His preoperative examination and investigations were unremarkable. Inhalational induction was done with sevoflurane and intravenous line secured with 24G cannula. After induction, i-gel TM of size 1 was introduced in single attempt. Anesthesia was maintained with sevoflurane, fentanyl, and atracurium. The lungs were ventilated with Jackson-Rees breathing circuit and the ventilation was ascertained by square wave capnography and adequate tidal volume. The second child was 12 kg, 3-year-old, and was scheduled to undergo MRI brain for a workup of delayed milestones. His Mallampati grading was class II. Intravenous line was already in place with 22G cannula. General anesthesia was induced with propofol, fentanyl, and atracurium. After induction of anesthesia, i-gel TM of size 2 was inserted. Nasogastric tube was not passed in the two patients.
Monitoring in both the patients was done by MRI compatible equipment that included continuous electrocardiography, pulse oximetry, and capnography. Hemodynamic parameters were stable throughout the procedure and ventilation adequate in both patients without any episode of desaturation. In both the cases, the procedure was completed uneventfully with a good image quality without artifacts. Neuromuscular blockade were reversed and i-gel was removed uneventfully after the children were fully awake.
Tracheal tubes or LMAs have been standard of care for the cases that need general anesthesia. The radiologists are usually familiar with the effect on imaging interpretations produced by the metallic spring in the pilot balloon or anatomic distortions induced by the LMAs, although there are some reports of misdiagnosis as a result of the LMAs. [1],[2] Any magnetic material can reduce the image quality depending on the quantity of magnetic material within the field, the pulse sequence that is used, and if the area of interest is in the region of the LMA. When the MRI is done in close proximity to LMA-Flexible, LMA-ProSeal, LMA-Fastrach or tracheal tube, is distortion of the image due to the wire component of the airway tube and heating. [3]
i-gel TM is a unique, single use, supraglottic airway management device. It is made from a medical grade thermoplastic elastomer and has been designed to create a noninflatable, anatomical seal of the pharyngeal, laryngeal, and perilaryngeal structures. This offers the potential for easier insertion, reduced tissue compression, and increased stability after insertion. [4] Avoidance of tracheal intubation in a child is less traumatic and the risk of adverse reflex reaction or laryngospasm during intubation or extubation is greatly reduced. [5] Although we used muscle relaxant to place the device, i-gel TM can be placed in spontaneously breathing patients, which may be safer during MRI. The presence of a gastric tube adds to the safety of supraglottic devices in such settings.
There are concerns regarding displacement of supraglottic devices during change in position of the patient. This is particularly true in MRI suites. With i-gel TM , it is possible to move the child in complete safety without getting the device displaced. It also means that intubation is not required as often. [6]
i-gel TM is a useful device for MRI in children. It offers many advantages such as ease of insertion, availability of all pediatric sizes, stability of device once in situ, avoidance of tracheal intubation, and has minimum adverse effects after removal of the device. Furthermore, the presence of gastric tube adds to safety of i-gel especially in spontaneously breathing patients. However, large sample-size studies are needed to confirm the usefulness of i-gel in MRI suite.
References | |  |
1. | SchiebleT, Patel A, Davidson M. Laryngeal mask airway (LMA) artifact resulting in MRI misdiagnosis. Pediatr Radiol 2008;38:328-30.  |
2. | Langton JA, Wilson I, Fell D. Use of laryngeal mask airway during magnetic resonance imaging. Anaesthesia 1992;47:532.  [PUBMED] |
3. | LMA TM Airway Inst ruction Manual. San Diego : L0 MA North America Inc.; 2005.  |
4. | Jolliffe L, Jackson I. Airway management in outpatient setting : N0 ew techniques and devices. Curr Opin Anaesthesiol 2008;21:719-22.  [PUBMED] [FULLTEXT] |
5. | Singh I, Gupta M, Tandon M. Comparison of clinical performance of i-gel TM with LMA- Proseal TM in elective surgeries. Indian J Anaesth 2009;53:302-5.  [PUBMED] |
6. | Bopp C, Carrenard G, Chauvin C, Schwaab C, Diemunsch P. The I-gel in paediatric surgery : I0 nitial series. American Society of Anesthesiologists (ASA) Annual Meeting; New Orleans, USA, October 17-21 2009 A 147.  |
This article has been cited by | 1 |
Airway Management in out of the Operating Room Environments |
|
| Shahehaz Elsamragy,Tarini Vats,Irene Osborn | | Current Anesthesiology Reports. 2020; | | [Pubmed] | [DOI] | | 2 |
A retrospective study of the impact of supraglottic airway devices on the appearance of neck masses in children undergoing serial magnetic resonance imaging |
|
| Allan F. Simpao,Vladislav Obsekov,Jorge A. Gálvez,Rebecca S. Isserman,Mohamed A. Rehman,Erin S. Schwartz | | Pediatric Anesthesia. 2018; | | [Pubmed] | [DOI] | | 3 |
Impact of airway management strategies on magnetic resonance image quality |
|
| F. E. Ucisik-Keser,T. L. Chi,Y. Hamid,A. Dinh,E. Chang,D. Z. Ferson,T. Asai | | British Journal of Anaesthesia. 2016; 117(suppl 1): i97 | | [Pubmed] | [DOI] | | 4 |
Update on Airway Devices |
|
| Lisa Sohn,Razan Nour,Narasimhan Jagannathan | | Current Anesthesiology Reports. 2015; | | [Pubmed] | [DOI] | | 5 |
Challenge to pediatric anatomical variation : Can we draw the ideal line on the pediatric I-gel? |
|
| Yukako Abukawa,Koichi Hiroki,Hiroko Iwakiri,Tomoko Fukada,Makoto Ozaki | | Journal of Anesthesia. 2015; | | [Pubmed] | [DOI] | | 6 |
Use of i-gel in magnetic resonance imaging |
|
| Kiranpreet Kaur,Mamta Bhardwaj,Prashant Kumar,Jatin Lal,Sanjay Johar,Sarla Hooda | | Acta Anaesthesiologica Taiwanica. 2014; | | [Pubmed] | [DOI] | |
|
 |
 |
|