LETTER TO EDITOR
Year : 2012  |  Volume : 28  |  Issue : 2  |  Page : 267-268

An unusual defect in the nitrous oxide pipeline


Department of Anaesthesia and Intensive Care, Postgraduate Institute of Medical Education and Research, Chandigarh, India

Date of Web Publication11-Apr-2012

Correspondence Address:
Amit Jain
Department of Anaesthesia and Intensive Care, Postgraduate Institute of Medical Education and Research, Chandigarh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9185.94921

Rights and Permissions

How to cite this article:
Jain A, Makkar JK, Sahni N, Solanki SL. An unusual defect in the nitrous oxide pipeline. J Anaesthesiol Clin Pharmacol 2012;28:267-8

How to cite this URL:
Jain A, Makkar JK, Sahni N, Solanki SL. An unusual defect in the nitrous oxide pipeline. J Anaesthesiol Clin Pharmacol [serial online] 2012 [cited 2021 Jan 24];28:267-8. Available from: https://www.joacp.org/text.asp?2012/28/2/267/94921

Sir,

We wish to report an unusual defect in nitrous oxide hose (Amvex, Richmond Hill, ON) that was detected while delivering anesthesia to a child. Routine checkup of all the equipment was done before starting the case in accordance with association guidelines and no leaks were detected. [1] A 3-year-old boy was scheduled for herniorrhaphy. Inhalational induction of anesthesia was started with oxygen (O 2 ), nitrous oxide (N 2 O), and sevoflurane, using an Ohmeda Aestiva/5 machine (Ohmeda, Helsinki, Finland) with a Jackson Rees breathing system. Sevoflurane was increased progressively by 2% every 3-4 breaths. Despite delivery of 8% sevoflurane for 2 mines, the child continued to struggle. Inadequate filling of the reservoir bag was observed. Circuit was checked for any accidental disconnections. Visual inspection of the N 2 O flow meter revealed a drop in the level of the bobbin from the initial setting of 3 to 0.8 l/min. An attempt to increase the N 2 O flow rate, by rotating the knob, failed to increase the level of the bobbin. A fall in the N 2 O pipeline pressure (on the anesthetic work station gauge) from 45 psig initially to 30 psig was also noted. Induction of anesthesia was achieved with sevoflurane in 100% O 2 at fresh gas flow rate of 4 l/min. Following induction, a quick search was made for possible cause of fall in N 2 O pressures. The manifold pressures were at 45 psig and the machine had been checked for any leak before starting the case.

A close inspection of the N 2 O hose demonstrated 18 cm long segment of fusiform dilatation [Figure 1] with a minute hole [Figure 2]. This portion of the hose was soft on texture and crepitations could be felt over it. Defect occurred possibly following the initial damage of internal nylon braided pipe resulting in the seepage of N 2 O (at intermediate pressures) between the external polyvinylchloride (PVC) coating and nylon braiding. The gas leaked between the two layers of pipe, before puncture of the PVC coating resulting in sudden drop in the N 2 O pipeline pressures occurred at the machine end. The most likely cause of the defect was the repeated wheeling of the anesthesia machine over the pipeline. Pipeline of N 2 O was changed later.
Figure 1: Nitrous oxide pipeline with arrow showing a dilated segment

Click here to view
Figure 2: Arrow pointing hole in the dilated portion of nitrous oxide pipeline

Click here to view


Such a damage to the N 2 O pipeline could present as sudden burst of pipeline and can even occur in an oxygen pipeline. The damage can be prevented by raising pipelines off the floor or by using gas columns/pendants/ceiling hose drops/articulating arms. Inspection of the anesthesia pipelines along with the test recommended for leaks [2] should be performed carefully and this should be added to the list of defects reported in the anesthesia gas pipeline systems. [3],[4],[5]

 
  References Top

1.AAGBI. Checklist for anaesthetic apparatus. London: Association of Great Britain and Ireland; 1997.  Back to cited text no. 1
    
2.Lecky JH. The mechanical aspects of anesthetic pollution control. Anesth Analg 1977;56:769-74.  Back to cited text no. 2
[PUBMED]  [FULLTEXT]  
3.Lacoumenta S, Hall GM. A burst oxygen pipeline. Anaesthesia 1983;38:596-7.  Back to cited text no. 3
[PUBMED]    
4.Howell RS. Piped medical gas and vacuum systems. Anaesthesia 1980;35:676-98.  Back to cited text no. 4
[PUBMED]    
5.Ewart IA. An unusual cause of gas pipeline failure. Anaesthesia 1990;45:498.  Back to cited text no. 5
[PUBMED]    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed1244    
    Printed56    
    Emailed0    
    PDF Downloaded240    
    Comments [Add]    

Recommend this journal