Intra-operative emergency airway management in a child with broken tracheal limb of Montgomery T-tube


Department of Anesthesiology, Pain Medicine and Critical Care, All India Institute of Medical Sciences, New Delhi, India

Date of Web Publication16-Jul-2018

Correspondence Address:
Kanil Ranjith Kumar
Department of Anesthesiology, Pain Medicine and Critical Care, All India Institute of Medical Sciences, New Delhi - 110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_256_17

Rights and Permissions

How to cite this article:
Singh T, Kumar KR, Parthiban M. Intra-operative emergency airway management in a child with broken tracheal limb of Montgomery T-tube. J Anaesthesiol Clin Pharmacol 2018;34:256-7

How to cite this URL:
Singh T, Kumar KR, Parthiban M. Intra-operative emergency airway management in a child with broken tracheal limb of Montgomery T-tube. J Anaesthesiol Clin Pharmacol [serial online] 2018 [cited 2021 Jan 18];34:256-7. Available from: https://www.joacp.org/text.asp?2018/34/2/256/236667



Madam,

Montgomery tube (MT) is placed after tracheal stenosis surgeries where it acts as a stent to prevent restenosis and also maintains the patency of the airway. It is a pliable, uncuffed, T-shaped tube made of silicone.[1] Various complications have been reported with MT in situ such as blockade by secretions, kinking of extra- and intra-tracheal lumen, migration, and inhalation. We report here an adverse event that occurred while replacing MT after surgery and emergency airway management to circumvent the situation.

A 12-years-old male child, an operated case of laryngotracheal reconstruction for subglottic stenosis, was posted for excision of granulations over glottis. He had 10-mm external diameter MT in situ following previous surgery. To preserve spontaneous ventilation, inhalational induction was performed with oxygen and sevoflurane by connecting MT to the circuit. After administering 60 μg of fentanyl, direct laryngoscopy was done to assess air leak from glottic aperture. The upper end of MT and glottic aperture was completely occluded by the granulation tissue. There was no leak on performing positive pressure ventilation. Atracurium was used for muscle relaxation. After removing granulation tissues over MT, a surgeon wanted to replace the MT with a tracheostomy tube (TT). After providing 100% oxygen for 3 min, the surgeon using a curved forceps attempted removal of MT. While doing so, the extratracheal limb of the MT broke off from the rest of the intratracheal limb [Figure 1]. This created an emergency situation with incompletely removed granulation obstructing at glottis and a broken MT obstructing at tracheostomy stoma. Meanwhile, as the surgeon struggled to remove the broken tube, saturation of the child dropped to 89%. Immediately, a 5-mm internal diameter uncuffed TT was inserted into the opening of broken MT that was visible through tracheostomy stoma. Ventilation was possible through the TT. Another attempt was made to remove the MT after keeping rigid bronchoscope ready. The intratracheal limb of MT was then removed successfully through the tracheostomy stoma. The child made an uneventful recovery and shifted to postanesthesia care unit.
Figure 1: Broken limbs of Montgomery T-tube after retrieval from the trachea

Click here to view


It is routine to replace MT with a TT or flexometallic tube before abolishing spontaneous breathing and proceeding to surgery. However, in this case, MT was adherent to granulation tissue, so tube change was attempted after releasing granulations. Airway compromise due to fracture of MT while attempted retrieval has not been reported previously. Restoring ventilation and oxygenation in such a scenario is of utmost importance. As the opening in broken MT was visible through tracheostomy stoma, a smaller size uncuffed TT was inserted into the MT by holding the upper limb of MT with forceps. Utmost care was taken to prevent distal migration of MT while inserting TT. Our plan B was to ventilate with facemask by occluding the tracheostomy stoma as the glottis was partially patent by now. On extreme situation, pushing the intratracheal MT into one of the bronchi was our last option.[2] Understanding the complexity and emergent nature of such situation and readiness with a plan of action are needed to prevent any undue complication.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Guha A, Mostafa SM, Kendall JB. The Montgomery T-tube: Anaesthetic problems and solutions. Br J Anaesth 2001;87:787-90.  Back to cited text no. 1
    
2.
Pawar DK. Dislodgement of bronchial foreign body during retrieval in children. Paediatr Anaesth 2000;10:333-5.  Back to cited text no. 2
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed1004    
    Printed19    
    Emailed0    
    PDF Downloaded0    
    Comments [Add]    

Recommend this journal