Unusual cause of airway obstruction after nasal intubation with a preformed endotracheal tube


Department of Anaesthesiology and Intensive Care, Adesh Institute of Medical Sciences and Research, Adesh University, Bathinda, Punjab, India

Date of Submission13-Mar-2019
Date of Acceptance24-Feb-2020
Date of Web Publication16-Oct-2020

Correspondence Address:
Prof. Minnu M Panditrao
Department of Anaesthesiology and Intensive Care, Adesh Institute of Medical Sciences and Research, Adesh University, Bathinda - 151 001, Punjab
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_71_19

Rights and Permissions

How to cite this article:
Bansal RB, Panditrao MM, Panditrao MM. Unusual cause of airway obstruction after nasal intubation with a preformed endotracheal tube. J Anaesthesiol Clin Pharmacol 2020;36:429-30

How to cite this URL:
Bansal RB, Panditrao MM, Panditrao MM. Unusual cause of airway obstruction after nasal intubation with a preformed endotracheal tube. J Anaesthesiol Clin Pharmacol [serial online] 2020 [cited 2020 Oct 28];36:429-30. Available from: https://www.joacp.org/text.asp?2020/36/3/429/298280

Dear Editor,

Nasal intubation with cuffed preformed endotracheal tubes (ETT) (RAE north facing) is a standard procedure for tonsillectomy surgery.[1] This procedure is carried out generally without complications. We report a case where after uncomplicated nasal intubation with a preformed RAE north-facing cuffed ETT, we were unable to ventilate the patient.

A 60-kg, 30-year-old ASA-I male was posted for tonsillectomy. As nasal intubation was planned, nasal vasoconstrictor (xylometazoline 0.1%) drops were instilled in both nostrils 30 min prior to induction of anaesthesia. Inside the operation theatre, standard monitoring was established. After preoxygenation, anaesthesia was induced with fentanyl 100 μg and propofol 140 mg. As bag-mask ventilation was easy, muscle relaxant atracurium 30 mg was given to facilitate intubation. After 3 min of mask ventilation, RAE preformed north facing 6.5-mm ETT (Portex) was inserted into the right nostril of patient. Laryngoscopy was done with size-3 Macintosh blade and laryngeal view was Cormack–Lehane grade-1. ETT was advanced into the trachea under direct vision using Magill forceps. Tube passage was without any resistance. ETT cuff was inflated with 10 mL of air and tube was fixed with elastic adhesive tape. Manual ventilation was initiated with a mixture of oxygen and nitrous oxide. Significant resistance in the reservoir bag and negligible chest expansion were noted. On auscultation, bilaterally breath sounds were not auscultated. All the connections from gas outlet to ETT were verified. Resistance was persistently high with no ETCO2 graph on the monitor. Left hypochondrium was auscultated for possible, oesophageal intubation, which was negative. Immediately suction catheter was passed through the ETT suspecting some mucous plug. But we failed to retrieve it. Laryngoscopy was performed again to confirm correct placement of ETT and it was confirmed. Treatment for suspected bronchospasm was given with injection hydrocortisone 100 mg, injection dexamethasone 8 mg, and injection deriphylline 2ml intravenously, but no improvement in ventilation occurred and SPO2 dropped to 60%. Suspecting ETT obstruction, patient's ETT was removed and bag-mask ventilation done, which was easy with visible chest expansion. Patient was reintubated with a new, the same sized (6.5-mm ID RAE North facing) tube. Now, the ventilation was without challenge. The remainder of the surgery and postoperative recovery were uneventful.

On examination of the removed ETT, a kink was seen in the region of ETT cuff [Figure 1] and [Figure 2] almost completely occluding the lumen of the tube. Although the exact cause of this problem remains unclear, we are hypothesizing two probable causes. First, as we were using 6.5-mm ID size ETT in an adult patient, there was ample space available in the tracheal lumen to allow the tube to fold on itself. The logic behind using a 6.5-mm ID size tube was extremely narrow external nares of the patient. Second, there could be a manufacturing defect leading to weakness of that part of the wall of ETT allowing it to kink.[1] It must be noted that as the RAE North facing tubes are to be negotiated nasally, they are generally made up of blue-coloured softer material making them more vulnerable for kinking.
Figure 1: Cuffed endotracheal tube (RAE) showing site of kinking right side view

Click here to view
Figure 2: Cuffed endotracheal tube (RAE) showing site of kinking (Bevel side view)

Click here to view


In anaesthesia practice, we often encounter difficulty in ventilating an endotracheally intubated patient. It could be due to many reasons like anaesthesia gas delivery malfunctions, breathing circuit obstruction (from common gas outlet to the end of ETT), poor pulmonary compliance, acute severe bronchospasm, tension pneumothorax, and endobronchial mass lesion. Thus, double-checking of the anaesthesia equipment and ETT before use and vigilant monitoring for ventilation are standards of anaesthesia practice. Acute endotracheal tube obstruction is a potentially life-threatening situation and rapid recognition and prompt action on any change in ventilatory parameters are essential. Common causes of ETT obstruction reported are mucous plug, blood clot,[2] or intratracheal kinking.[3] Structural defects[4] of ETT are also not very rare and can happen in any part of the ETT. Defects in ETT cuff, over- inflated cuff herniation, intraluminal tracheal obstruction, intraluminal plastic films,[5] and plastic membrane of tube connector causing severe airway obstruction have been reported.[6]

One must be vigilant and act promptly to diagnose and treat the cause of airway obstruction which can present at any time during the course of anaesthesia.

We conclude that when encounter airway obstruction in an otherwise “normal everything,” “inadvertent and hidden” intratracheal kinking also must be kept in mind.

Authors contribution

The second author conceptualized the main theme of the article. The first author did most of the literature search and writing, while the corresponding author did most of the editing and revising and the second author did the final corrections in the preparation of manuscript. In fact, all the authors contributed almost equally.

Declarations

The authors want to confirm that all the authors had full access to all the data in the manuscript and the corresponding author takes the full and final responsibility for the decision to submit for publication. The manuscript has been read and approved by all the authors. The requirements for authorship have been met and each author believes that the manuscript represents honest work.

The first author has taken the responsibility of the integrity of the work as a whole from inception to published article and has been designated as “guarantor."

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Lee YW, Lee TS, Chan KC, Sun WZ, Lu CW. Intratracheal kinking of endotracheal tube. Can J Anesth 2003;50:311-5.  Back to cited text no. 1
    
2.
Park C, Kim H, Yum K. Acute obstruction of an endotracheal tube: A case report. Anaesth Proq 2004;51:62-4.  Back to cited text no. 2
    
3.
Leissner KB, Ortega R, Bodzin AS, Sekhar P, Stanley GD. Kinking of an endotracheal tube within the trachea: A rare cause of endotracheal obstruction. J Clin Anaesth 2007;19:75-6.  Back to cited text no. 3
    
4.
Chua WL, Ng AS. A defective endotracheal tube. Singapore Med J 2002;43:476-8.  Back to cited text no. 4
    
5.
Sofi K, EI-Gammal K. Endotracheal tube defect: Hidden causes of airway obstruction. Saudi J Anaesth 2010;4:108-10.  Back to cited text no. 5
    
6.
Baldemir R, Akcaboy Y, Akcaboy ZN, Gogus N. Endotracheal tube obstruction: A manufacturing defect. Turk J Anaesthiol Reanim 2015;43:62-4.  Back to cited text no. 6
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed76    
    Printed0    
    Emailed0    
    PDF Downloaded14    
    Comments [Add]    

Recommend this journal